Product Description
Product Advantages:
1. High temperature resistance; capable of long-term operation in a 260°C environment
2. High voltage resistance; internal fluoroplastic probe withstands high voltage AC3000V
3. Compact size; basically applicable to various confined spaces
4. Fast heat conduction; faster response speed and higher accuracy
Electrical Performance:
1. Insulation Resistance: 500VDC > 100MΩ
2. Dielectric Strength Test: 1500VAC 60s ≤ 1mA
3. Drop Test: Free fall from a height of 3 meters, no visible damage
4. Flammability Rating: UL94 V-0
5. Vibration Test: 10-500Hz 3G
6. Temperature Measurement Range: -50°C ~ 260°C
7. Temperature Measurement Accuracy: ±(0.3 + 0.05|t|)°C, where "t" represents the actual temperature
Measurement Current and Self-Heating:
The power supply current heats the platinum thin-film sensor. The resulting temperature measurement error is calculated using the following formula:
Δt = P × SP, Power loss = I²RS, Self-heating coefficient unit is K/mW.
The datasheet specifies the self-heating coefficient for each product.
Self-heating is determined by the thermal contact between the platinum thin-film sensor and the surrounding medium.
If the heat dissipation to the environment is highly efficient, a higher measured current can be obtained.
There is no specified lower limit for the measured current of platinum thin-film sensors. It largely depends on the application.
We recommend:
For 100 Ohm: Maximum 1 mA